Current Output 2 sub-function

In this sub-function you set the current output range of the transmitter for
temperature values.

ENT

ENT

ENT

ENT

ENT

ENT

ENT

o

c

o

c

o

c

o

c

2

2

3

3

4

4

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

1

cur.2

SETUP

HOLD

1. Select the “CUR.2” sub-function, then press the ENT key.

2. Selecting the output type: press the

▲ or ▼ key to select the desired

output type: 0-20 mA or 4-20 mA. Press the ENT key to confirm your
selection.

3. Setting the temperature value at which the transmitter output will be

4 mA (or 0 mA): press the

▲ or ▼ key to set the temperature value

(setting range:

-9.9 to 115°C or 15 to 237 °F) to be equivalent

to 4 mA (or 0 mA). Press

the ENT key to confirm your setting.

4. Setting the temperature value at which the transmitter output will be

20 mA:

press the

▲ or ▼ key to set the temperature value (setting

range: -0.1 to 125°C or 35 to 257 °F) to be equivalent

to 20 mA. Press the

ENT key to confirm your setting.

Continue with Setup mode procedures, or return to Measurement mode by
pressing the

▲ and ▼ keys simultaneously (escape).