Wash relay (WASH) sub-function

In this sub-function you set the parameters for the wash relay.

1

2

2

3

3

4

5

ENT

ENT

ENT

ENT

ENT

ENT

ENT

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

off

1. Select the “WASH” sub-function, then press the ENT key.

2. Enabling/disabling wash function: press the

▲ or ▼ key to enable

(WASH ON) or disable (WASH OFF) wash function. Press the ENT key to
confirm your selection.

3. Selecting the relay status condition: press the

▲ or ▼ key to choose

desired relay status (de-energized = “DEEN” or energized = “EN”). Press
the ENT key
to confirm your selection.

4. Setting the wash interval in hours: Press the

▲ or ▼ key to set the

desired wash interval (setting range: 0.1 to 199.9 hours). Press the ENT
key
to confirm your setting.

5. Setting the wash duration in seconds: Press the

▲ or ▼ key to set the

desired wash duration (setting range: 1 to 1999 seconds). Press the ENT
key
to confirm your setting.

Note: During wash cycle the controller is set to HOLD. For safety
reasons, the HOLD function is activated 5 seconds prior and
deactivated 10 seconds after the wash cycle.

Continue with Setup mode procedures, or return to Measurement mode by
pressing the

▲ and ▼ keys simultaneously (escape).