Thermo Fisher Scientific Alpha pH 2000 User Manual

Page 38

Advertising
background image

– “PH SY” = pH measurement with symmetrical input

– “PH ASY” = pH measurement with asymmetrical input

– “ORP % SY” = ORP-% measurement with symmetrical input

– “ORP % ASY” = ORP-% measurement with asymmetrical input

– “ORP mV SY” = ORP-mV measurement with symmetrical input

– “ORP mV ASY” = ORP-mV measurement with asymmetrical input

Press the ENT key to confirm your selection.

5. Setting the alarm delay time in seconds: press the

▲ or ▼ key to set

the alarm delay time (setting range: 0 to 1999 seconds). Press the ENT
key
to confirm your setting.

Note: With this parameter you set the delay before the alarm relay
activates when the set point has been exceeded.

6. Selecting the alarm contact type: press the

▲ or ▼ key to select the

alarm contact type:

– “STDY” = steady contact

– “FLET” = fleeting (single pulse) contact

Press the ENT key to confirm your selection.

Note: With this parameter you select whether the alarm contact will
operate as a steady contact or a fleeting (single pulse) contact. Pulse
contact closing time is 1 second.

7. Selecting the sensor type: press the

▲ or ▼ key to select the sensor

type which is connected to the controller:

– “GLAS” = glass electrode

– “ANTY” = antimony electrode

Press the ENT key to confirm your selection.

Note: This parameter is not available in ORP mode.

Note: After each changeover, the factory calibration data for zero point
and slope are overwritten. Be sure to recalibrate (see section 5).