TECHNICAL SPECIFICATIONS

8.1

General Specification

Alpha pH 2000 Controller / Transmitter
pH Range

-2.00 to 16.00 pH

Resolution & Accuracy

0.01 pH & ± 0.01 pH

mV Range

0 to 100% or -1000 to 1000 mV

Resolution & Accuracy

0.1% or 1 mV / ± 1 mV

Temperature

-9.9 to 125

o

C (15.0 to 257.0

o

F)

Resolution & Accuracy

0.1 & ± 0.5

o

C (± 1.0

o

F)

Sensor

Pt 1000 or Pt 100 (jumper selectable)

Temperature Compensation

Automatic (± 10

o

C / ± 18

o

F offset adjustment) /

Manual


Set-point And Controller Functions
Function

Limit / Proportional Control / Proportional Integral
(Pulse Length or Pulse Frequency)

Integral time

0 to 999.9 minutes

Adjustable period with pulse
length controller

0.5 to 20 sec

Adjustable period with pulse
frequency controller

60 to 120 pulses/min

Pickup / Dropout Delay

0 to 1999 seconds

Wash Cycle

0.1 to 199.9 hours

Wash Duration

1 to 1999 seconds

Switching pH Hysteresis

0.1 to 1 pH

Switching ORP Hysteresis

1 to 10.0 % / 10 to 100 mV

Contact Outputs, Controller

1 SPDT; 3 SPST relays

Switching
Voltage/Current/Power

Max 250 VAC / Max 3A / Max 600 VA


Alarm Functions

Function (switchable)

Latching / pulse

Pickup delay

0 to 1999 seconds

Switching
Voltage/Current/Power

Max 250 VAC / Max 3A / Max 600 VA


Display
LCD

UV coat, backlit 14 segments display with symbols
for status information

Backlight

On/Off selection with four level of brightness
control