6 current output 1 sub-function, Urrent, Utput – Thermo Fisher Scientific Alpha pH 2000 User Manual

Page 34: Function, Cur.1, Current output 1 sub-function

Advertising
background image

10. Setting the proportional range: press the

▲ or ▼ key to set the

proportional range (setting range: 10 to 500%). Press the ENT key to
confirm your setting.

11. Setting the pulse frequency: press the

▲ or ▼ key to set the pulse

frequency (setting range: 60 to 120 pulses per minute). Press the ENT
key
to confirm your setting.

12. Setting the integral action time: press the

▲ or ▼ key to set the

integral action time (setting range: 0.0 to 999.9 minutes). Press the ENT
key
to confirm your setting.

Note: If the integral action time is set to ‘0.0’, then the unit functions as a
‘P’ Controller.

Continue with Setup mode procedures, or return to Measurement mode by
pressing the

▲ and ▼ keys simultaneously (escape).

6.6

Current Output 1 sub-function

In this sub-function you set the current output range of the transmitter for pH
or mV values.

2

2

3

3

4

4

5

5

SETUP

HOLD

PH

SETUP

HOLD

PH

SETUP

HOLD

PH

SETUP

HOLD

PH

SETUP

HOLD

PH

SETUP

HOLD

SETUP

HOLD

PH

SETUP

HOLD

1

cur.1

SETUP

HOLD

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

1. Select the “CUR.1” sub-function, then press the ENT key.

2. Selecting the output type:

press the

▲ or ▼ key to select the desired

output type: 0-20 mA or 4-20 mA. Press the ENT key to confirm your
selection.

3. Setting the pH or ORP value at which the transmitter output will be 4

mA (or 0 mA):

press the

▲ or ▼ key to set the pH or ORP value to be

equivalent to 4 mA or 0 mA, respectively. Press the ENT key to confirm
your setting.

4. Setting the pH or ORP value at which the transmitter output will be 20

mA: press the

▲ or ▼ key to set the pH or ORP value to be equivalent

to

20 mA. Press the ENT key to confirm your setting.

5. Selecting out of range current: Press the

▲ or ▼ key to switch “CU22”

on or off. If “CU22” is on, the output signal will jump to 22 mA when the
measured pH and ORP value is below or above the values set in points 3
& 4 above. Press the ENT key to confirm your setting.

Note: the “CU22” setting (on/off) of current output 1 applies to current

output 2 as well.

Continue with Setup mode procedures, or return to Measurement mode by
pressing the

▲ and ▼keys simultaneously (escape).