Resistivity Setup

PC6000, CON6000, PC6500 and PCD6500 meters

Technology Made Easy ...

To Set Alarm Limits
This option allows you to set alarm limits for the resistivity measuring mode. If the resistivity value
of the measurement is outside of the boundaries set by the minimum and maximum limits,
audible and visual warnings will let you know that your sample measurement was outside of the
set limits.

The Alarm Limit can be set from 3.000 Ohm to 100.000 M Ohm.

To set Alarm Limits

1. Touch ON or OFF to set the status of the alarm of resistivity mode
2. Touch the Low box and use numeric keypad to enter the new limit values.
3. Touch Enter on the keypad to accept this limit and return to the resistivity Setup screen.
4. Touch Ohm, KOhm or MOhm to select resistivity units of low alarm value.
5. Touch the High box and use numeric keypad to enter the new limit values.
6. Touch Enter on the keypad to accept this limit and return to the resistivity Setup screen.
7. Touch Ohm, KOhm or MOhm to select resistivity units of high alarm value.




To Set Print Criteria
This screen allows you to select which criteria are printed with the measurement when you print
the data or send it to a computer. The status of the current print criteria is displayed on the
screen. The criteria option is active if ON appears to the right of the option. It is inactive if OFF
appears to the right of the option. Any active criteria will be printed on demand.

The Resistivity Print Criteria is set the same way as previously described in Conductivity Setup.
See Conductivity Setup: Set Print Criteria’ section for instructions.