Thermo Fisher Scientific CyberScan 6000 Series Meters v.2 User Manual

Page 56

Advertising
background image

141

DO Operation: Temperature Standardization

DO6000 and PCD6500 meters

Technology Made Easy ...

To Standardize Temperature of the Meter

1. Touch Standardize on the DO measure screen to access the standardized screen.
2. Make sure DO probe with built ATC is attached to meter.
3. Immerse your DO probe with built ATC probe into a solution of known temperature, such

as a temperature bath, for a few minutes while temperature stabilizes.

4. Touch Temp Std to access temperature standardization.
5. The Temperature Standardization screen appears. Check the current temperature

displayed with that of the solution of known temperature. Touch the Standardization
Temperature
box and use numeric keypad to enter the current temperature. Press Enter
to confirm value.

6. Touch OK to confirm Standardization Temperature and return to the DO Measure screen.

Touch Reset to reset and enter a new standardization temperature. Touch Cancel to
cancel temperature standardization and return to the DO Standardization screen.

Note: The meter will not allow entered value to exceed ±5°C of the ATC probe value. If
entered value exceeds ±5°C an error window will appear and you will have to re-enter
value or cancel out of the temperature screen. This might indicate that you have a faulty
ATC probe.