Thermo Fisher Scientific CyberScan 6000 Series Meters v.2 User Manual
Page 57

142
DO Operation: DO Measurement
DO6000 and PCD6500 meters
Technology Made Easy ...
The measurement mode is used to measure Dissolved Oxygen (DO) in %Sat, mg/L or mbar. The
DO measure screen provides a readout from the DO probe attached to the meter. Before
measuring DO, you will need to standardize your probe: see ‘DO Operation: Standardizing’
section for details. It is good practice to standardize the probe daily for best results.
If the Auto Read is active, the meter will lock (freeze) onto a reading when it is stable. The meter
will not change from this reading until the Measure button is pushed, at which time a new reading
is taken. If the Auto Read is not active, the meter will continuously monitor the DO and the
reading will indicate any change in the DO level. If the STABLE indicator is turned on, the
STABLE indicator appears on the screen when the reading meets the chosen stability criteria.
NOTE: Prior to sample measurement, verify the salinity of the sample and enter that salinity value
using the setup menu (see ‘DO Setup: Set Salinity / Set Alarm Limits’ section
for details).
To Measure DO with Auto Read ON
1. Immerse rinsed DO probe into
the sample solution, constantly
stirring the solution.
2. Touch Measure.
3. The meter will continuously
measure the sample until the
stability criteria is met. At this
time the reading will freeze and
the STABLE* indicator appears.
4. Record the reading once the
measurement has become
stable.
5. To take another reading, repeat
steps 1 to 4.
* Please note that STABLE will only appear if you
have completed a successful two point calibration.
To Measure DO with Auto Read OFF
1. Immerse rinsed DO probe into
the sample solution, constantly
stirring the solution.
2. The meter will continuously
measure the sample until the stability criteria is met. At this time the STABLE indicator
appears.
3. Record the reading once the measurement has become stable.
4. To take another reading, repeat steps 1 to 3.
NOTES: You can access other functions of the meter with the remaining buttons on the measure
screen.
• Touching Setup will access the Setup screen.
• At anytime, you can touch Mode to access another mode of operation including: BOD, OUR and
SOUR.