Thermo Fisher Scientific CyberScan 6000 Series Meters v.2 User Manual

Page 62

Advertising
background image

147

DO Operation: BOD Measurement

DO6000 and PCD6500 meters

Technology Made Easy ...

To measure BOD of the Seed:

1. Insert DO probe into a BOD bottle filled with your seed.
2. Allow probe to stabilize for 30 minutes, constantly stirring the solution.
3. Touch BOD on the BOD measurement screen to enter BOD Calculation Parameter

Screen. Screen show current DO reading.

4. Touch the SEED to designate BOD measurement as seed.


Initial DO measurement for Seed BOD calculation
The Bottle ID text entry box allows you to enter the bottle ID of the sample for new reading. The
Select Existing ID scroll box is in the selection of Add new sample/Seed option for new seed
reading.

To enter Bottle ID:

1. Touch the Bottle ID box.
2. Use the alphanumeric keypad to enter the desired Bottle ID. The BS key allows you to

remove characters that are incorrectly entered.

3. Touch Enter to accept current Bottle ID and return to the BOD Calculation Parameter

Screen.

4. After entering the Bottle ID, the following

text entry boxes will appear and require
more information regarding the current
sample:

a. Description
b. Seed Volume
c. Bottle Volume

5. After entering the above parameters for a

new sample reading, touch Apply BOD
Detail
button to accept the initial reading
for a seed used to calculate BOD.

NOTE: The stored data is displayed in the Available Sample ID text box in tree view. By
selecting the + sign for a particular Bottle ID, all the stored parameters for that particular Bottle ID
will display in the Available Sample ID text box. Select the - sign to close the tree view of the
sample.

Advertising