Thermo Fisher Scientific CyberScan 6000 Series Meters v.2 User Manual

Page 96

Advertising
background image

181

Appendix: Replatinization

PC6000, CON6000, PC6500 and PCD6500 meters

Technology Made Easy ...

Replatinization is the process of replacing the platinum on the surfaces of the 2 cell conductivity
probes that may flake or wear off over time. The platinum on the surface of the probe is used to
increase the surface area of the measuring surface resulting in decreased polarization error.
Replatinization is a relatively quick procedure to perform, taking no more than 5 minutes.
Although replatinization is not a long process, the replatinization solution is costly.

To Perform Replatinization

1. Access the Replatinization mode from the Mode screen. In any measuring mode, touch

Mode to access the Mode screen.

2. Touch Replat. to enter the Replatinization mode.
3. Immerse your 2 cell conductivity probe into the replatinization solution. The immersion

depth should be enough to cover the vent holes on the probe.

4. Touch Start on the display screen to begin the process of replatinization. At this time a

countdown timer will appear on the display screen showing the time elapse and time
remaining in the replatinization procedure.

Note: At anytime after you have initiated the replatinization:

You can touch Stop to stop the procedure

OR

You can touch Exit to cancel the replatinization the procedure and return to the Mode screen.

4-cell conductivity probes do not require replatinization.

Advertising