AMPLIFIER MOUNTING CONFIGURATION OPTIONS

Meter Mount Configuration

The meter mount configuration has the amplifier mounted directly on the detector . This compact, self-contained

configuration minimizes installation wiring .

Remote Mount Configuration

The remote mount configuration places the amplifier at a location away from the fluid flow and detector . This is necessary in

situations where process fluid temperature or the environment exceeds amplifier ratings .
The detector and amplifier are connected by wires, run through conduit, between junction boxes on the detector and the

amplifier . The distance between the detector junction box and amplifier junction box can be up to 100 feet (30 meters) . A

remote mounting bracket is supplied .

Submersible Option

If you are installing the meter in a vault, you should order the remote amplifier option . You must not install the amplifier inside

a vault . We also recommend ordering the remote meter package with the submersible option (NEMA 6P) . This will eliminate

any potential problems resulting from humidity or temporary flooding in the vault .

OTEE:

N

NEMA 6P enclosures are constructed for indoor or outdoor use to provide protection against access to hazardous

parts; to provide a degree of protection against ingress of solid foreign objects and water (hose directed water and

the entry of water during prolonged submersion at a limited depth); that provide an additional level of protection

against corrosion and that will be undamaged by the external formation of ice on the enclosure .

WIRING

Wiring Safety

AT INSTALLATION, BE SURE TO COMPLY WITH THE FOLLOWING REQUIREMENTS:

• Disconnect power to the unit before attempting any connection or service to the unit .
• Do not bundle or route signal lines with power lines .
• Keep all lines as short as possible .
• Use twisted pair shielded wire for all output wiring .
• Observe all applicable, local electrical codes .

Opening the M5000 Cover

The M5000 amplifier's design lets you open the cover

without completely removing it .
Follow these steps:
1 . Completely remove the top two screws from the

amplifier using a blade/slotted screwdriver .

2 . Loosen both of the bottom screws so that the round

head of each screw clears the top face of the cover .

3 . Pull the cover down to the open position .

Figure 17: Remove Two Screws

Figure 18: Open the Cover