Inputs/outputs menu – Badger Meter M-Series M5000 User Manual

Page 25

Advertising
background image

Inputs/Outputs Menu

Inputs/Outputs

Flow Simulation

S

Flow Simulation provides output simulation based on a percentage of the full scale flow . Simulation

will not accumulate the totalizers . The range of simulation includes –100…100% of the full scale flow .
The Flow Simulation Parameter lets you set the range of simulation in increments of 50 (OFF, 0, 50, 100,

-50, -100) . The factory default is OFF .

Digital Input

S

• Digital input lets you reset totalizers or interrupt flow measurement .
• Input switching is provided by applying an external voltage of 3…35V DC .
• Use a “normally open” contact for operating .

Digital Outputs

(continued on next

page)

Pulse/Unit

S

This parameter lets you set how many pulses per unit of measure will be

transmitted to remote applications . For example, assuming the unit of measure

is gallons:

• Setting the Pulses/Unit to 1 (standard setting) will transmit 1 pulse

every gallon

• Setting the Pulses/Unit to 0 .01 will transmit 1 pulse every 100 gallons

You must configure pulses/unit if the function of the selected output is to be

forward or reverse .

This parameter must be considered with the Pulse Width and Full Scale Flow

parameters . The maximum pulse frequency is 100 Hz . The frequency is correlated

with the flow rate . Violation of output frequency limits will generate a PULSE_

OVERLOAD_WARNING .

Width

S

This parameter establishes the ON duration of the transmitted pulse . The

configurable range is 0…500 ms . The factory default is 0 ms .

• Non-zero pulse width configuration – the OFF duration of the transmitted

pulse is dependent on flow rate . The OFF duration is to be at least the

configured ON duration . At full scale flow, the ON duration equals the Off

duration . The maximum configurable output frequency is limited to 100 Hz .

• The duty cycle of the transmitted pulse is at 50% of the output frequencies

greater than 1 Hz .

This parameter must be considered with the Pulses/Unit and Full Scale Flow

parameters . The maximum pulse frequency is 100 Hz . The frequency is correlated

with the flow rate . Violation of output frequency limits will generate a PULSE_

OVERLOAD_WARNING .

Power

Line Frequency

S

This parameter provides measuring immunity to industrial noise from a power

supply feed .
This parameter should be set to your nominal line frequency, typically 50 Hz .

Using the m5000 main menu programming options

MAG-UM-00219-EN-03

Page 25

August 2014