Errors & Warnings

What You See

Why It Happened

How to Fix It

MEASURE_TIMEOUT

The board may be damaged .

Contact Badger Meter Technical Support .

COMMON_MODE_VOLTAGE_

OVERLOAD

Common mode voltage is smaller than
–2 .0V or larger than +4 .1V .

Make sure the meter is properly grounded .

Empty detector .

Switch Empty Pipe detection ON .

Dirty electrodes .

Clean the electrodes .

EMPTY_PIPE_WARNING

Measured impedance between the Empty Pipe
electrode and the Ground exceeded the set value .

• Make sure the pipe is filled .

• Check Empty Pipe threshold . It should be

60000 Ω (corresponds with 20 µS/cm) .

Empty detector .

PULSE_OVERLOAD_WARNING

Overflow occurred on the flow output .

Lower the pulse number .

FLOW_OVERLOAD_WARNING

Flow exceeded the Full Scale of more than 100% .

Set the flow range properly .

LOW_POWER_WARNING

Battery voltage is smaller than 3 .0V .

Consider replacing the battery .

EEPROM_ERROR

Configuration file is missing .

Replace the board .

CONFIG_ERROR

Configuration file is corrupted .

Replace the board .

PREAMPLIFIER_OVERLOAD

Input voltage exceeded the limits .

Maximum polarization is ± 227 mV; maximum
power line noise is 10 .6 mV; maximum useful
signal is 10 .7 mV .

Empty detector .

Switch Empty Pipe detection ON .

Dirty electrodes .

Clean the electrodes .

COIL ERROR

Coil/sensor is not connected .

• Make sure wiring is properly connected to the

amplifier .

• If that does not clear the alarm, contact

Badger Meter Technical Support .

Coil shorted .
Issue with coil current .

Troubleshooting