Range 2 Calibration Information:
4. When in P4.1, press ▲ to enter P4.2 to

review the last calibration concentration.

If range 2 is not yet calibrated, “---” will

appear on the LCD.

5. When in P4.1 or P4.2, press

SET for

less than one second to confirm and

return to P4.0.

6. When in P4.0, press ▲ to advance to

the next setup function or press

SET

for more than 2 seconds to return to

Measurement Mode.

Note…

P4.1 and P4.2 Modes allow review of the

calibration information but are not used

during the calibration procedure.

Selecting the Measurement Range (rAn)
There are two measurement ranges.

The pen defaults to auto-ranging:

Salinity

Range 1

0 ~ 10.00 ppt

Range 2

10.1 ~ 70.0 ppt