Sper Scientific 850037 Large Display Conductivity Pen User Manual

Page 41

Advertising
background image

41

9. Stir the probe gently to create a uniform

sample. Wait about 15 minutes for the

reading to stabilize.

10. The unit of measure

will flash on the LCD to

indicate that the pen is in

Measurement Mode. When

the reading is stable, the

unit of measure will stop

flashing.

Hold Function
1. Press

HLD to freeze the

current readings on the

display. “Hold” will appear

on the LCD.

2. Press

HLD again to release

the hold function.

Set the temperature coefficient to

the right value. The meter is factory

set to 2.1% per °C and this normally

can provide good results. See page

9 to set the temperature coefficient

if needed.

Select the normalization temperature.

The meter is factory set to 25 °C. If

you need to normalize the meter to

20 °C, see page 9

Rinse the probe with deionized or

distilled water before use to remove

any impurities adhering to the

electrode. If the meter is not

used for a long time, please soak the

probe more than 30 minutes to clear

up the lazy effect of probe.

Dip the probe into the sample. Make

sure there are no air bubbles trapped

in the slot of the probe. To remove

air bubbles, give the probe a gentle

stir. Making sure the electrode tip

is submerged when you stir it.

Stir the probe gently in the sample

to create a homogenous sample.

Allow a few seconds to reach temp.

equilibrium. (Wait about 15 minutes

can get a stable reading)

for the details.

The unit of measurement

will flash on the LCD to

indicate the meter is in

measurement mode. When

the reading is stable, the unit

will stop flashing.

8351

4.

5.

6.

7.

8.

9.

5

“ "

Press

to freeze

current readings. The text

"Hold" will appear on the

LCD. P ress " " again to

release.

8351

In 8361 measurement mode, press

" " to switch to conductivity or TDS

mode. The units for conductivity are

uS or mS. The units for TDS are ppm

or ppt.

Cond.

TDS

6

In

TDS measurement mode, you can

press " " more than 2 seconds to

select manual ranging. (See details in

page11)

10.

TDS Measurement:

Set the TDS conversion factor to right

value. The factory default is 0.50.

To change the TDS factors, refer

Appendix B in page 24.

11.

Salinity Measurement:

12.

8371 is designed for salt measurement.

When in SALT mode, Built-in NaCL

conductivity to TDS conversion factor,

Temp. Coefficient and Normalization

Temperature are all fixed.

You may follow up steps 3-10 to

proceed the salt measurement.

See page 12

for the detail