The default value of the reference

temperature is set at 25°C. To normalize to

a different value, see page 33. Determine

the calibration standard value for your

reference temperature before resetting

the reference temperature. Contact the

solution distributor for the calibration

standard value of the solution.

Selecting the Calibration Schedule
Calibration is necessary and should be

performed regularly. Soak the probe

for 15 minutes before calibration or

measurement to minimize drift. If the pen

is used in the mid ranges, calibrate the

unit monthly.

If the measurement is performed at

extreme temperatures or within the

concentration listed below, calibrate

the unit weekly.

Model

Concentration

Conductivity Pen

< 100 uS

> 2 mS