Sper Scientific 850037 Large Display Conductivity Pen User Manual

Page 36

Advertising
background image

36

7. Press

CAL for more than 2 seconds to

begin calibration. The conductivity value

of the solution will flash on the LCD.

8. Press ▲ or ▼ to adjust the value on

the primary display to match the value

of the standard buffer. You can adjust

the conductivity reading ± 30% from the

detected value. If the detected value

and the standard value differ by more

than ± 30%, clean or replace the pen.

Example:

Standard: 10 uS; Detected value: 19 uS

Adjustable range: ± 5.7 uS (19*30%)

Under the above situation, the values

differ by over 30%.

Note…

If the standard value is over the measuring

range or 10% less, the displayed value will

be equal to the range limit or 10% of the

range limit.

Example 1:

Standard: 22 uS; Detected value: 19 uS

Adjustable range: ± 5.7 uS (19*30%)