Sper Scientific 850037 Large Display Conductivity Pen User Manual

Page 55

Advertising
background image

55

21

Error code

FigA

FigB

FigC

FigD

t

Many air bubbles adhered!

To reduce the air bubbles, stir the

electrode completely and better to

dip the electrode into solution at

oblique angle. To dip the electrode

at vertical angle is easy to create

many air bubbles.

After soaking the electrode in

solution for 15~30 minutes, inspect

the electrode carefully to make sure

no bubbles adhere. If air bubbles

still exist, tap the bottom of the

container gently and stir the electrode

to remove the air bubbles. If above

method are not working, remove

the electrode out of solution and

blow at the electrode to remove the

air bubbles.

To remove the air bubbles:
1. Stir the probe.

2. Dip the probe into the solution at

an oblique angle (at a slant, not

perpendicular to solution). Dipping

the probe perpendicularly will cause

air bubbles to form.

3. Soak the probe in solution for 15-30

minutes. Inspect the probe carefully for

air bubbles.

4. If air bubbles remain on the probe, tap

the bottom of the container gently and

stir the probe.

5. If the preceding methods fail, remove

the probe from the solution and blow on

the probe to remove the air bubbles.