ERROR CODES
Salinity
- - - The pen is in manual ranging 1 but

the salinity measurement is higher

than 10.0 ppt; Press ▲ for more

than 2 seconds to change to manual

ranging 2 or auto-ranging mode.

E03 The salt value is over the range limit

(70.0 ppt) or the pen is damaged;

Put the pen in standard buffer

solution (the buffer concentration

must be lower than the range limit).

If E03 still appears, send the pen to

Sper Scientific for repair.

E04 Error in measuring original data

(damaged temperature sensor,

temperature out of specifications

or salinity error). E04 will disappear

when the original error is resolved.