Although the values differ less than 30%,

the 22 uS is still over the range limit. The

maximum input value is 19.99 uS.

Example 2:

Standard: 1.6 mS; Detected value: 2.1 mS

Adjustable range: ± 0.63 mS (2.1*30%)

Although the measured value differs less

than 30%, the 1.6 mS is still less than the

10% range limit (1.99). The maximum

input value is 1.99 mS.

9. When “CAL” stops flashing on the LCD,

press

SET for less than 1 second to

confirm the value. The pen will return to

Measurement Mode. To exit Calibration

Mode without saving, press

SET for

more than 2 seconds. This allows you

to retain the pen’s previous calibration

data for the current range as you

proceed. If “CAL” flashes on the LCD

continually, check that the calibration

solution is stable and that the input

value (Step 8) is correct.

10. Repeat Steps 1-9 for other ranges

as needed.