43

ERROR CODES
Conductivity
- - -

The pen is in manual ranging 1 but

the conductivity measurement is

higher than 1999 uS; Press ▲ for

more than 2 seconds to change to

manual ranging 2 or auto-ranging

mode.

E03 The conductivity value is over

the range limit (19.99 mS) or the

pen is damaged; Put the pen in

standard buffer solution (the buffer

concentration must be lower than

the range limit). If E03 still appears,

send the pen to Sper Scientific for

repair.

E04 Error in measuring original data

(damaged temperature sensor or

temperature out of specifications)

results in this conductivity value

error; Refer to the temperature

error codes (page 44). E04 will

disappear when the temperature

error is resolved.