2 using display/keypad commands, 3 using led commands – Vaisala HMP260 User Manual

Page 47

Advertising
background image

HMP260 SERIES

HMP260-U017en-1.5

Operating Manual

1997-11-25

41

•

If you want to see how the sensor stabilizes to the humidity in the
calibrator, enter c <cr> instead of the first reference:

RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...

•

Place the sensor head into the calibration hole of the NaCl bottle in
the humidity calibrator.

•

Wait for 10 minutes.

•

Press any key and enter the second point value and press <cr>.

RH : xx.x Ref2 ? yy.y <cr>

•

The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.

6.1.2.2

Using display/keypad commands

Disconnect the security lock jumper!

•

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

•

Place the sensor head into the calibration hole of the LiCl bottle in the
humidity calibrator.

•

Wait for 10 minutes.

•

Select Cali in the main menu and then RH cal; select Not changed
and then two-point calibration RH 2 point cal. Change the first point
reading with the arrow keys and press ENT.

•

Place the sensor head into the calibration hole of the NaCl bottle in
the humidity calibrator.

•

Wait for 10 minutes.

•

If necessary, change the second point reading with the arrow keys and
press ENT.

6.1.2.3

Using LED commands

Disconnect the security lock jumper!

•

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

•

Place the sensor head into the calibration hole of the LiCl bottle in the
humidity calibrator.