CALIBRATION COMMANDS

CRH

Relative humidity calibration

Disconnect the security lock jumper!

CRH <cr>

With this command the transmitters can be calibrated against a reference. A one-point
calibration can be done against an accurate transfer standard in the field and a two-point
calibration using saturated salt solutions in controlled conditions. A two-point calibration
is performed as follows:

CRH <cr>

RH : 12.00 Ref1 ? 11.3 <cr>

Press any key when ready ...

RH : 76.00 Ref2 ? 75.5 <cr>

In one-point offset correction, the Ref2 prompt is acknowledged with <cr>:

CRH <cr>

RH : 12.80 Ref1 ? 11.3 <cr>

Press any key when ready ...

RH : 75.50 Ref2 ? <cr>

If the stabilization of the sensor to the humidity in the calibrator needs to be monitored,
the measurement output can be repeated by giving command c<cr> at Ref1 and Ref2:

CRH <cr>

RH : 12.00

Ref1 ? c <cr>

RH : 11.70

Ref1 ? c <cr>

RH : 11.50

Ref1 ? 11.3 <cr>

Press any key when ready ...

RH : 76.00

Ref2 ? 75.5 <cr>