HMP260 SERIES

HMP260-U017en-1.5

Operating Manual

1997-11-25

43

•

Press any key and enter the second point value and press <cr>.

RH : xx.x Ref2 ? yy.y <cr>

•

The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.

6.1.3.2

Using display commands

Disconnect the security lock jumper!

•

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

•

Place the sensor head into the calibration hole of the LiCl bottle in the
humidity calibrator.

•

Wait for 10 minutes.

•

Select Cali in the main menu and then RH cal; select Sensor
changed. Change the first point reading with the arrow keys and
press ENT.

•

Place the sensor head into the calibration hole of the NaCl bottle in
the humidity calibrator.

•

Wait for 10 minutes.

•

If necessary, change the second point reading with the arrow keys and
press ENT.

6.1.3.3

Using LED commands

Disconnect the security lock jumper!

•

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

•

Place the sensor head into the calibration hole of the LiCl bottle in the
humidity calibrator.

•

Wait for 10 minutes.

•

Connect an ammeter/voltmeter to the analogue outputs (connector
X2). Give command

l¡¡¡. At the first calibration point the LED

on the left flashes; adjust the first point with the arrow switches to the
value given in the calibration table (Chapter 6.1.4) and press ENT
switch.

•

Place the sensor head into the calibration hole of the NaCl bottle in
the humidity calibrator.

•

Wait for 10 minutes.