Appendix 1: Serial commands

or pressure is set to zero using XPRES. After this the pressure stored using command
PRES is valid again.

CDATE

Entering calibration date

CDATE xxxxxx <cr>

xxxxxx =

calibration date (000101...991231)

When the latest calibration date has to kept in memory, it is entered as follows:

CDATE 940506 <cr>

If the command is given without the date, the transmitter outputs the latest calibration
already in memory.

CDATE <cr>

940420

The date can be given in any format; however, the maximum number of numbers is six.

DATE

Setting the date

DATE <cr>

For example, to enter a new date:

DATE <cr>

Current date is 1993-01-30

Enter new date (yyyy-mm-dd) : 1993-06-12 <cr>

When the current date is asked, the new date is passed with <cr>.

TIME

Setting the time

TIME <cr>

For example, to enter a new time:

TIME <cr>

Current time is 01:35:54

Enter new time (hh:mm:ss) : 13:25:56 <cr>

When the current time is asked, the new time is passed with <cr>.