background image

HMP260 SERIES

HMP260-U017en-1.5

Appendix 1: Serial commands

4

OTHERS

ITEST

Testing the analogue outputs

ITEST <cr>

or

ITEST a b <cr>

a

=

current/voltage of channel 1

b

=

current/voltage of channel 2

The operation of the analogue outputs can be tested by forcing the outputs to given
values which can then be measured with a current/voltage meter from the analogue
outputs. The response to ITEST command gives six outputs/parameters. Only the first
two are relevant; they show the channel current or voltage in mA or V. The other four
figures contain information for service purposes only.

Examples:

reading the channel outputs and parameters

>itest <cr>

1.9438 2.3483 1.00694 10.64634 1.97374 2.17665

>

forcing outputs 0.5 V and 4 V to channels 1 and 2

>itest 0.5 4 <cr>

0.5000 4.0000 1.00694 10.62970 1.23336 3.01722

>

releasing the forced control and reading the outputs

>itest <cr>

1.9427 2.3392 1.00731 10.62428 1.97157 2.16978

>

MTIM

Setting the measurement integration time

MTIM nnn <cr>

nnn

=

number of cycles measured (4...255)

By lengthening the measurement integration time any stray changes in the output can be
filtered out: the transmitter calculates the average of a number of measurement cycles
defined by the user. The command can be given in two ways: