Appendix 1: Serial commands

FCRH

Relative humidity calibration after sensor change

Disconnect the security lock jumper!

FCRH <cr>

After humidity sensor change the transmitter must be calibrated using this command and
the calibration must be done at two reference points. The calibration is performed as
follows:

FCRH <cr>

RH : 1.90

Ref1 ?

11.3 <cr>

Press any key when ready ...

RH : 76.30

Ref2 ?

74.9 <cr>

The command can also be divided into two commands, so the computer can be used
for other purposes while waiting for the sensor to stabilize to the higher humidity.

FCRH 1 <cr>

RH : 1.90

Ref1 ?

11.3 <cr>

FCRH 2 <cr> ...

RH : 76.30

Ref2 ?

74.9 <cr>

If the stabilization of the sensor to the humidity in the calibrator needs to be monitored,
the measurement output can be repeated by giving command c<cr> at Ref1 and Ref2:

FCRH <cr>

RH : 12.00

Ref1 ? c <cr>

RH : 11.70

Ref1 ? c <cr>

RH : 11.50

Ref1 ? 11.3 <cr>

Press any key when ready ...

RH : 76.00

Ref2 ? 75.5 <cr>

CT

Temperature calibration

Disconnect the security lock jumper!

CT <cr>

With this command the transmitters can be calibrated against an accurate reference,
such as a Pt 100 simulator. A two-point calibration is performed as follows:

CT <cr>

T : 0.80 Ref1 ?

0.0 <cr>

Press any key when ready ...

T : 56.20

Ref2 ?

55.0 <cr>