Appendix 1: Serial commands
FCRH
Relative humidity calibration after sensor change
Disconnect the security lock jumper!
FCRH <cr>
After humidity sensor change the transmitter must be calibrated using this command and
the calibration must be done at two reference points. The calibration is performed as
follows:
FCRH <cr>
RH : 1.90
Ref1 ?
11.3 <cr>
Press any key when ready ...
RH : 76.30
Ref2 ?
74.9 <cr>
The command can also be divided into two commands, so the computer can be used
for other purposes while waiting for the sensor to stabilize to the higher humidity.
FCRH 1 <cr>
RH : 1.90
Ref1 ?
11.3 <cr>
FCRH 2 <cr> ...
RH : 76.30
Ref2 ?
74.9 <cr>
If the stabilization of the sensor to the humidity in the calibrator needs to be monitored,
the measurement output can be repeated by giving command c<cr> at Ref1 and Ref2:
FCRH <cr>
RH : 12.00
Ref1 ? c <cr>
RH : 11.70
Ref1 ? c <cr>
RH : 11.50
Ref1 ? 11.3 <cr>
Press any key when ready ...
RH : 76.00
Ref2 ? 75.5 <cr>
CT
Temperature calibration
Disconnect the security lock jumper!
CT <cr>
With this command the transmitters can be calibrated against an accurate reference,
such as a Pt 100 simulator. A two-point calibration is performed as follows:
CT <cr>
T : 0.80 Ref1 ?
0.0 <cr>
Press any key when ready ...
T : 56.20
Ref2 ?
55.0 <cr>