HMP260 SERIES

HMP260-U017en-1.5

Operating Manual

1997-11-25

45

6.2.1

One point offset correction

6.2.1.1

Using serial commands

Disconnect the security lock jumper!

•

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

•

Check the transmitter against the reference.

•

Give command CT <cr> and enter the first point value and press
<cr>:

CT <cr>
'T : xx.x Ref1 ? yy.y <cr>
Press any key when ready

•

If you want to see how the sensor stabilizes to the reference
temperature, enter c <cr> instead of the first reference:

T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...

•

After giving the correct temperature value (Ref1) and pressing <cr>
press any key and press <cr>.

6.2.1.2

Using display/keypad commands

Disconnect the security lock jumper!

•

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

•

Check the transmitter against the reference.

•

Select Cali in the main menu and then T cal; select two-point
calibration T 2 point cal. Change the first point reading with the
arrow keys and press ENT.

•

Select

T 1 point cal, change the reading to correspond to the

reference and press ENT.