HMP260 SERIES
HMP260-U017en-1.5
Appendix 1: Serial commands
The output scales can also be given directly with the ASCL command.
ASCL aaa.a bbb.b ccc.c ddd.d <cr>
aaa.a
=
lower limit of channel 1
bbb.b
=
upper limit of channel 1
ccc.c
=
lower limit of channel 2
ddd.d
=
upper limit of channel 2
For example, when relative humidity is scaled to 0...100 %RH on channel 1 and
temperature to -40...+100
°
C on channel 2:
ASCL 0 100 -40 100 <cr>
Ch1
(RH)
lo
0.000 %RH
Ch1
(RH)
hi
100.000 %RH
Ch2
(T ) lo
-40.000 'C
Ch2
(T ) hi
100.000 'C