Appendix 1: Serial commands

ADDR

Setting the transmitter address

ADDR aa <cr>

aa

=

address (0...99)

The address is used when more than one transmitter is connected to one serial bus. The
ADDR command makes it possible to communicate with one transmitter at a time in
POLL mode.

For example, transmitter is given address 99

ADDR <cr>

Address : 2 ? 99 <cr>

When asking the current address, no address number is given:

ADDR <cr>

Address : 2 ? <cr>

RESET

Resetting the transmitter

RESET <cr>

Resets the transmitter. All settings that have been changed stay in the memory even after
reset or power failure.

3.1

Operation modes

SMODE

Setting the serial interface

SMODE xxxx<cr>

xxxx

=

STOP, RUN or POLL

In STOP mode:

measurements output only by command, all commands can be used

In RUN mode:

outputting automatically, only command S can be used

In POLL mode:

measurements output only with command SEND. When in POLL
mode, the output mode is changed as follows:

Advertising