The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.

6.2.2.2

Using display/keypad commands

Disconnect the security lock jumper!

•

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

•

Check the transmitter against the reference.

•

Select Cali in the main menu and then T cal; select two-point
calibration T 2 point cal. Change the first point reading with the
arrow keys and press ENT.

•

Change the temperature and check the transmitter again against the
reference.

•

Check that the reading corresponds to the reading of the reference
instrument. If not, adjust the second point

•

If necessary, change the second point reading with the arrow keys and
press ENT.

6.2.2.3

Using LED commands

Disconnect the security lock jumper!

•

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

•

Check the transmitter against the reference.

•

Connect an ammeter/voltmeter to the analogue outputs (connector
X2). Give command

¡¡l¡. At the first calibration point the LED

on the left flashes; adjust the first point (offset) with the arrow
switches to the same reading with the reference and press ENT
switch.

•

Change the temperature and check the transmitter again against the
reference.

•

Check that the reading corresponds to the reading of the reference
instrument. If not, adjust the second point.

•

If necessary, adjust with the arrow switches to the correct value and
press ENT. At the second calibration point the second LED from the
left flashes.