Data recording, Selecting data recording quantities, Dsel – Vaisala MMT330 User Manual

Page 110: Nd data recording on pa

Advertising
background image

User's Guide _______________________________________________________________________

110 __________________________________________________________________ M210764EN-F

NOTE

When using the RS-485 interface with a 2-wire connection, always
disable echo. When using a RS-232, RS-422/485 4-wire connection,
LAN, or WLAN, you can enable or disable it as you wish.

Data Recording

Data recording function is always on and collects data automatically into
the memory of the device. If the optional data logger module is installed,
the transmitter uses it automatically. Recorded data does not disappear
from the memory when the power is switched off. Collected data can be
observed in a form of a graph in the graphical view of the display or it
can be listed out by using the serial line or MI70 Link program.

Selecting Data Recording Quantities

If the device is provided with the optional display, the recorded quantities
are always those selected for the display. Up to three quantities can be
recorded at a time. See section Changing Quantities and Units on page 97
for instructions on how to select the display quantities with the keypad.

DSEL

Use the serial line command DSEL to select the quantities to be recorded
if the transmitter is not equipped with display/keypad. The MMT330
measures the following quantities:

- Water activity (aw)

- Temperature (T) (metric unit: ºC, non-metric unit:ºF)

DSEL [xxx]<cr>

where

xxx = Data recording quantity.

Example:

>dsel aw t
aw T
>

Enter the command without parameters and press ENTER to display the
current recording parameters.