Configuration registers, Exception status outputs, Table 37 – Vaisala MMT330 User Manual

Page 167: Configuration parameter registers, Table 38, Configuration flag registers, Table 39, Mmt330 exception status outputs, Exception status outputs on pa

Advertising
background image

Appendix A _________________________________________________________ Modbus Reference

VAISALA _______________________________________________________________________ 167

Configuration Registers

Configuration parameter registers are used to configure the measurement.
Writing out-of-range values is silently ignored.

Table 37

Configuration Parameter Registers

Name

Float

Integer

Valid Range

Pressure setting default
value

0769…0770

1025 (×1)

0…9999 hPa

Pressure setting
temporary value

0771…0772

1026 (×1)

0…9999 hPa

Oil-A

0777…0778

1029 (×0.01)

±99999.99

Oil-B

0779…0780

1030 (×0.0001) ±99.9999


If you modify the pressure setting repeatedly, modify the temporary
value at register 0771...0772 (1026) instead of the permanent setting at
register 0769…0770 (1025). Set the temporary value to zero to return to
the default pressure setting.

Non-metric units are not available on Modbus. If they are needed,
calculate the conversion outside the transmitter.

Configuration flags are used to select some basic options of the
instrument and to manually start the sensor operations.

Table 38

Configuration Flag Registers

Name

Address

Description

Standard filtering on/off

1281

1 = Filtering on

Extended filtering on/off

1282

1 = Extended filtering on


Flags 1281 and 1282 are mutually exclusive.

Some of the configuration registers might not have an effect depending
on the instrument model and configuration.

Other configuration settings are not usually needed and are not available
using the Modbus interface. The service cable must be used to change
other configuration settings if needed.

Exception Status Outputs

Exception status outputs (read with function code 07) tell the summary of
the MMT330 transmitter status as described in Table 39 below.

Table 39

MMT330 Exception Status Outputs

Output

Name

Description

0 (0x01)

Fault status

1 = No errors

1 (0x02)

Online status

1 = Online data available

Advertising