Diagnostic modbus counters, Viewing counters using display/keypad, Viewing counters using service port – Vaisala MMT330 User Manual

Page 133: Figure 66, Modbus counters

Advertising
background image

Chapter 5 ___________________________________________________________________ Modbus

VAISALA _______________________________________________________________________ 133

For a description of the available settings, see section Wireless

LAN Configuration on page 85.

8.

Modbus configuration is now complete. Reset or power cycle the
transmitter to enable the Modbus mode, and proceed with the
installation of the transmitter.

Diagnostic Modbus Counters

MMT330 has diagnostic counters that can be used to pinpoint Modbus
problems. The counters are always active when the Modbus protocol is
enabled.

Viewing Counters Using Display/Keypad

You can use the display/keypad option to view and clear the counters.
Enter the Main Menu and navigate to

System ► Diagnostics

► MODBUS Counters.

1101-037

Figure 66

Modbus Counters

Viewing Counters Using Service Port

Use the MODBUS command to view the counters:

MODBUS<cr>

Example:

>modbus
Bus messages : 0
Bus comm. error: 0
Bus exceptions : 0
Slave messages : 0
Slave no resp. : 0
Last message :
>

In addition to the diagnostic counters, the last message (received or
transmitted) is shown in hexadecimal format. If the last message was a
broadcast message, the suppressed response message is shown.

Advertising