Vaisala MMT330 User Manual

Page 148

Advertising
background image

User's Guide _______________________________________________________________________

148 __________________________________________________________________ M210764EN-F

Example (one-point adjustment):

>ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>

5.

OK indicates that the calibration has succeeded. Give the
calibration information (date and text) to the transmitter's memory;
see the serial commands CTEXT and CDATE.

6.

Press the ADJ button on the motherboard to close the adjustment
mode.

7.

Take the probe out of the reference conditions and replace the
filter.

Advertising