Formatting serial line message, Ftime and fdate, General settings – Vaisala MMT330 User Manual

Page 97: Changing quantities and units, On changing quantities and units on pa, General settings on pa, 97 f

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 97

Formatting Serial Line Message

NOTE

Instead of using the FTIME and FDATE commands described in this
section, you can use the FORM command with modifiers TIME and
DATE. See section FORM on page 98.

FTIME and FDATE

FTIME and FDATE commands will enable/disable output of time and
date to the serial line. To add time to R and SEND outputs enter:

FTIME [x]<cr>

To add date to R and SEND outputs enter:

FDATE [x]<cr>

where

x

=

ON or OFF

Example:

>send
aw= 0.277 T= 23.5 'C
>ftime on
Form. time : ON
>send
00:16:07 aw= 0.277 T= 23.5 'C
>fdate on
Form. date : ON
>send
2000-01-01 00:16:15 aw= 0.277 T= 23.5 'C
>

General Settings

Changing Quantities and Units

To change quantities and units use serial commands or the optional
display/keypad. The MMT330 measures the following quantities:

- Water activity (aw)

- Temperature (T) (metric unit: ºC, non-metric unit:ºF)

NOTE

Only the quantities selected when ordering the device can be selected as
display output quantities.