Error states, Figure 68, Error indicator and error message – Vaisala MMT330 User Manual

Page 137

Advertising
background image

Chapter 6 _______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 137

Error States

In error state the affected quantities are not measured and the output is
shown as follows:

- Analog channel outputs 0 mA or 0 V (you can use the serial line

command AERR or display/keypad to change this fault indication
value, see section Analog Output Fault Indication Setting on page
118.)

- Serial port outputs stars "***" instead of measurement data

- Display shows "----" instead of measurement data

- Cover LED is blinking

- Optional display: error indicator is lit.

1103-068

Figure 68 Error Indicator and Error Message

The following number refers to Figure 68 above:

1

=

Error indicator

The error indicator disappears when the error state is over and you have
checked the error message. Press the INFO button to display the error
message.

You can also check the error message via the serial interface by using the
command ERRS. In case of constant error, please contact Vaisala; see
section Technical Support on page 139.

Advertising