Analog output quantities, On analog output quantities on pa – Vaisala MMT330 User Manual

Page 115

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 115

Example: 0 ... 5 V voltage output selected for channel 1 and 4...20 mA
selected for channel 2.

OFF ON

Selection

1

2

Voltage output selected

3

4

5

6

0 ... 5 V selected

7

8

1

Current output selected

2

3

4

4 ... 20 mA selected

5

6

7

8

NOTE

If you have customized the error output setting (AERR), check that the
set error values are still valid after changing the output mode/range. See
section Analog Output Fault Indication Setting on page 118.

Analog Output Quantities

Use the display/keypad to change and scale the analog output quantities.

1.

Press any of the arrow buttons to open the Main Menu.

2.

Select Interfaces

by pressing the ► arrow button.

3.

Select Analog outputs

by pressing the ► arrow button.

4.

Select Output 1/2/3

by pressing the ► arrow button.

5.

Select Quantity

by pressing the ▲▼ arrow buttons. Confirm your

selection by pressing CHANGE.

6.

Select the quantity by using the arrow buttons. Press SELECT to
confirm your selection.

7.

Select Scale, lower limit,

by pressing the ▲▼arrow buttons. Press

SET to confirm your selection. Press OK to confirm your setting.

8.

Select the upper limit

by pressing the ▲▼arrow buttons. Use the

arrow buttons to set the upper limit value. Press SET to confirm
your selection. Press OK to confirm your setting.

9.

Press EXIT to return to the basic display.