146 __________________________________________________________________ M210764EN-F

RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>

11.

OK

indicates that the adjustment has succeeded and the new

calibration coefficients are calculated and stored.Enter the
adjustment information (date and text) to the memory of the
transmitter, see the commands CTEXT and CDATE.

12. Press the ADJ button on the motherboard to close the adjustment

mode.

13. Take the probe out of the reference conditions and replace the

filter.

Relative Humidity Adjustment after Sensor
Change

Using Display/Keypad

When using the optional display/keypad, follow the instructions in
section Using Display/Keypad on page 144 but select Adj. for new RH
sensor (
instead of 1-point/ 2-point adjustment).

Using Serial Line

After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.

FCRH

Example:


>fcrh
RH

:

1.82 1. ref

?

0

Press any key when ready...

RH

:

74.22

2. ref

? 75

OK
>

The

OK

indicates that the calibration has succeeded.