PROGRAMMING MODE LOCK-OUT

Programming Mode is locked-out, as long as activated

(maintained action). A security code can be configured to

allow programming access during lock-out.

ZERO (TARE) DISPLAY

The Zero (Tare) Display provides a way to zero the Input Display value at

various input levels, causing future relative input display readings to be offset.

This function is useful in weighing applications where the container or material

on the scale should not be included in the next measurement value. When

activated (momentary action), rESEt flashes and the Display is set to zero. At

the same time, the Display value (that was on the display before the Zero

Display) is subtracted from the Display Tare Value and is automatically stored

as the new Display Tare Value. If another Zero (tare) Display is performed, the

display again changes to zero and the Display Tare Value shifts accordingly.

RESET TARE VALUE

The Reset Tare provides a way to zero the Display Tare (offset) value,

eliminating the Tare (offset) from the relative display. When activated

(momentary action), rESEt flashes and the Display Tare value is set to zero.

Following a Reset Tare, the Input display (relative) value will match the Gross

(absolute).

RELATIVE/GROSS (ABSOLUTE) VALUE

This function will switch the Input Display between Relative and Gross

(Absolute) value. The Relative is a net value that includes the Display Tare

(Offset)Value. The Input Display will show the Relative unless switched by this

function. The Gross is an absolute value (based on Input (Analog) Module dSP

and INP entries) without the Display Tare (Offset) Value. The Gross value is

selected as long as the user input is activated (maintained action) or at the

transition of the function key (momentary action). When the user input is

released, or the function key is pressed again, the input display switches back to

Relative value. GrOSS (gross) or rEL (relative) is momentarily displayed at

transition to indicate which value is being displayed.

HOLD DISPLAY

The active display is held but all other meter functions

continue as long as activated (maintained action).

HOLD ALL FUNCTIONS

The meter disables processing the input, holds all

display contents, and locks the state of all outputs as long

as activated (maintained action). The serial port continues

data transfer.

SYNCHRONIZE METER READING

The meter suspends all functions as long as activated

(maintained action). When the user input is released, the

meter synchronizes the restart of the A/D converter input

sampling with other processes or timing events.

STORE BATCH READING IN TOTALIZER

The Input Display value is added (batched) to the Totalizer when activated

(momentary action) and the display flashes bAtCh. The Totalizer retains a

running sum of each batch operation until the Totalizer is reset. When this

function is selected, the normal operation of the Totalizer is overridden and only

batched Input Display values accumulate in the Totalizer.

SELECT TOTALIZER DISPLAY

The Totalizer appears on Line 2 as long as activated

(maintained action). When the user input is released, the

previously selected display is returned. The

D or P keys

override and disable the active user input. The Totalizer continues to function

including associated outputs independent of the selected display.

RESET TOTALIZER

When activated (momentary action), rESEt flashes and the Totalizer resets

to zero. The Totalizer then continues to operate as it is configured. This selection

functions independent of the selected display.

RESET AND ENABLE TOTALIZER

When activated (momentary action), rESEt flashes

and the Totalizer resets to zero. The Totalizer continues to

operate while active (maintained action). When the user

input is released, the Totalizer stops and holds its value. This selection functions

independent of the selected display.

ENABLE TOTALIZER

The Totalizer continues to operate while active

(maintained action). When the user input is released, the

Totalizer stops and holds its value. This selection functions

independent of the selected display.

SELECT MAXIMUM DISPLAY

The Maximum display appears on Line 2 as long as

activated (maintained). When the user input is released,

the previously selected display is returned. The

D or P

keys override and disable the active user input. The Maximum continues to

function independent of the selected display.

RESET MAXIMUM DISPLAY

When activated (momentary action), rESEt flashes and the Maximum resets

to the present Input Display value. The Maximum function then continues from

that value. This selection functions independent of the selected display.

SELECT MINIMUM DISPLAY

The Minimum display appears on Line 2 as long as

activated (maintained). When the user input is released,

the previously selected display is returned. The

D or P

keys override and disable the active user input. The Minimum continues to

function independent of the selected display.

USEr-n

PLOC

FNC

USEr-n

rEL

FNC

Fn

rEL

FNC

USEr-n

r-tArE

FNC

Fn

r-tArE

FNC

USEr-n

d-rEL

FNC

Fn

d-rEL

FNC

USEr-n

d-HLd

FNC

USEr-n

A-HLd

FNC

USEr-n

SYNC

FNC

USEr-n

bAt

FNC

Fn

bAt

FNC

USEr-n

d-tot

FNC

USEr-n

r-tot1

FNC

Fn

r-tot

FNC

USEr-n

r-tot2

FNC

USEr-n

E-tot

FNC

USEr-n

d-HI

FNC

USEr-n

r-HI

FNC

Fn

r-HI

FNC

USEr-n

d-LO

FNC

Advertising