Code, Reset, P2-s – Red Lion PAX2S User Manual

Page 28: Factry ), Actory, Ervice, Perations

Advertising
background image

28

f

aCTOry

s

erviCe

O

peraTiOns

(FACtrY)

FACTORY SERVICE CODE

0-250

Enter the Service Code for the desired operation.

RESTORE FACTORY DEFAULTS

Use the ! and @ keys to display COdE 66 and press

P. The meter will flash

rESEt

and then return to COdE 50. Press the

P key to return to Display Mode.

This will overwrite all user settings with the factory settings. The only exception

is the User Mnemonics which retain their programmed values (see Code 69).

RESTORE FACTORY DEFAULTS (w/Units Mnemonics)

Same as Code 66, except the User Mnemonics are also returned to the factory

default settings (blank).

MODEL AND CODE VERSION

The meter will briefly display the model (P2S) on Line 1, and the current

firmware version (UEr x.xx) on Line 2, and then return to COdE 50.

METER CALIBRATION

NO INPUt AnLOut

The meter has been fully calibrated at the factory. Scaling to convert the input

signal to a desired display value is performed in Input Parameters. If the meter

appears to be indicating incorrectly or inaccurately, refer to Troubleshooting

before attempting to calibrate the meter. When recalibration is required

(generally every 2 years), it should only be performed by qualified technicians

using appropriate equipment. Calibration does not change any user programmed

parameters. However, it will affect the accuracy of the input signal and the

values previously stored using the Apply (APPLY) Scaling Style.

Preparation for Voltage Input Calibration

Warning: Input Calibration of this meter requires a signal source

capable of producing a signal greater than or equal to the range

being calibrated with an accuracy of 0.01% or better.

Before starting, verify that the Input Range Jumper is set for the range to be

calibrated. Verify that the precision signal source is connected and ready. Allow

a 30 minute warm-up period before calibrating the meter. Selecting NO at any

calibration step, will cause the unit to maintain the existing calibration

parameters for that step. Selecting YES and pressing the

P key will cause the unit

to store new calibration settings for the range selected. Pressing

D at any time

will exit programming mode, but any range that has been calibrated will

maintain the new settings.

Input Calibration Procedure

1. After entering CodE 48, in Factory Service Operations, select the input

(0.020v or 0.20v) to be calibrated.

2. Press the

P key until the desired range along with 2ER is indicated on Line 1

of the meter.

3. Apply the zero input limit of the range indicated on Line 1 of the meter.

4. Press ! to select YES.

5. Press

P. Display will indicate ---- on Line 2 as the unit reads and stores the

new calibration parameter.

6. Display will indicate the desired range along with FUL on Line 1 of the meter.

7. Apply the signal level indicated on Line 1 of the meter.

8. Press ! to select YES.

9. Press

P. Display will indicate ---- on Line 2 as the unit reads and stores the

new calibration parameter.

10. Repeat Preparation and Calibration Procedure for the other Input Range if

calibration for the other range is desired.

Analog Output Card Calibration

Before starting, verify that a precision meter with an accuracy of 0.05% or

better (voltmeter for voltage output and/or current meter for current output) is

connected and ready. Using the chart below, step through the five selections to

be calibrated. At each prompt, use the PAX2S ! and @ keys to adjust the

output so that the external meter display matches the selection being calibrated.

When the external reading matches, or if the range is not being calibrated, press

the

P key to advance to the next range. When all the desired ranges have been

calibrated, exit programming mode and remove the external meters.

DISPLAY

EXTERNAL METER

ACTION

0.000A

0.00 mA

Adjust if necessary, press

P

0.004A

4.00 mA

Adjust if necessary, press

P

0.020A

20.00 mA

Adjust if necessary, press

P

0.0v

0.00 V

Adjust if necessary, press

P

10.0v

10.00 V

Adjust if necessary, press

P

COdE

5O

FCS

COdE

66

FCS

rESEt

COdE

50

FCS

P

COdE

69

FCS

rESEt

COdE

50

FCS

P

COdE

51

FCS

P2-S

V Er

x.xx

FCS

COdE

50

FCS

P

COdE

48

FCS

CAL

NO

FCS

P