5 two-point slope only (option 3), Program – Campbell Scientific CR3000 Micrologger User Manual

Page 166

Advertising
background image

Section 7. Installation

166 

 

 

7.8.1.5.5 Two-Point Slope Only (Option 3)

Some measurement applications do not require determination of offset. Wave
form analysis, for example, may only require relative data to characterize change.

Case: A soil-water sensor is to be used to detect a pulse of water moving through
soil. To adjust the sensitivity of the sensor, two soil samples, with volumetric
water contents of 10% and 35%, will provide two known points.

The following procedure sets the sensitivity of a simulated soil water-content
sensor.

1. CRBasic example FieldCal Multiplier-Only Demonstration Program

(p. 166)

to

the CR3000.

2. To simulate the soil-water sensor, place a jumper wire between channels VX1

and SE6 (3L).

3. Simulate deployment-calibration conditions (output @ 10% = 175 mV, output

@ 35% = 700 mV) in two stages.

a. Set

variable

mV to 175. Set variable KnownWC to 10.0.

b.

Start the calibration by setting variable CalMode = 1.

c. When

CalMode increments to 3, set variable mV to 700. Set variable

KnownWC to 35.

d.

Resume the calibration by setting variable CalMode = 4

4. When variable CalMode increments to 6, the calibration is complete.

Calibrated multiplier is 0.0476.

CRBasic Example 30. FieldCal() Multiplier‐Only Demonstration Program 

'Jumper VX1 to SE6(3L) to simulate a sensor

Public

mV

'Excitation mV Output

Public

KnownWC

'Known Water Content

Public

CalMode

'Calibration Trigger

Public

Multiplier

'Sensitivity

Public

Offset

'Offset (Starts at zero, not changed)

Public

RelH2OContent

'Relative Water Content

'Data Storage Output of Calibration Data ─ stored whenever a calibration occurs

DataTable

(CalHist,NewFieldCal,200)

SampleFieldCal

EndTable

BeginProg

Multiplier = 1
Offset = 0
KnownWC = 0

LoadFieldCal

(true)

'Load the CAL File, if possible

Advertising