2 single-ended measurement reference, 3 ground potential differences, 1 soil temperature thermocouple – Campbell Scientific CR3000 Micrologger User Manual

Page 95: 2 external signal conditioner

Advertising
background image

Section 7. Installation

95

 

7.5.2 Single-Ended Measurement Reference

Low-level, single-ended voltage measurements are sensitive to ground potential
fluctuations. The grounding scheme in the CR3000 has been designed to eliminate
ground potential fluctuations due to changing return currents from 12V, SW12V,
5V, and C1C8 terminals. This is accomplished by utilizing separate signal
grounds (

) and power grounds (G). To take advantage of this design, observe

the following grounding rule:

Note Always connect a device ground next to the active terminal associated with
that ground. Several ground wires can be connected to the same ground terminal.

Examples:

• Connect grounds associated with 5V, 12V, and C1C8 terminals to G

terminals.

• Connect excitation grounds to the closest ( ) terminal on the excitation

terminal block.

• Connect the low side of single-ended sensors to the nearest ( ) terminal on

the analog input terminal blocks.

• Connect shield wires to the nearest ( ) terminal on the analog input terminal

blocks.

If offset problems occur because of shield or ground leads with large current flow,
tying the problem leads into the (

) terminals next to the excitation and pulse-

counter channels should help. Problem leads can also be tied directly to the
ground lug to minimize induced single-ended offset voltages.

7.5.3 Ground Potential Differences

Because a single-ended measurement is referenced to CR3000 ground, any
difference in ground potential between the sensor and the CR3000 will result in a
measurement error. Differential measurements MUST be used when the input
ground is known to be at a different ground potential from CR3000 ground.

Ground potential differences are a common problem when measuring full-bridge
sensors (strain gages, pressure transducers, etc), and when measuring
thermocouples in soil.

7.5.3.1 Soil Temperature Thermocouple

If the measuring junction of a copper-constantan thermocouple is not insulated
when in soil or water, and the potential of earth ground is, for example, 1 mV
greater at the sensor than at the point where the CR3000 is grounded, the
measured voltage is 1 mV greater than the thermocouple output, which equates to
approximately 25°C higher than actual.

7.5.3.2 External Signal Conditioner

External signal conditioners, an infrared gas analyzer (IRGA) is an example, are
frequently used to make measurements and send analog information to the
CR3000. These instruments are often powered by the same ac line source as the

Advertising