Typical performance for isotope ratio measurements is given in delta notation. For example, the δ

13

C for CO

2

is given

by:

13

1 1000

s

VPDB

R

C

R

δ

=

− ×

where R

s

is the ratio of the isotopomer concentrations measured by the TGA100 (

13

CO

2

/

12

CO

2

) and R

VPDB

is the

standard isotope ratio (

13

C/

12

C). δ

13

C is reported in parts per thousand (per mil or ‰). The 10 Hz noise is the square

root of the Allan variance with no averaging. The calibrated noise assumes a typical sampling scenario: two air sample
intakes and two calibration samples measured in a 1 minute cycle. It is given as the standard deviation of the calibrated
air sample measurements.

Table 2. Typical Isotope Ratio Measurement Noise

Gas

Isotope Ratio Wavenumber (cm

-1

) 10 Hz Noise (‰) Calibrated Noise (‰)

δ

13

C 2293.881,

2294.481 0.5

0.1

Carbon

Dioxide

δ

18

O 2308.225,

2308.416 2.5

0.5

δ

18

O 1500.546,

1501.188

2

0.5

Water

δD 1501.813,

1501.846 10

2.5

1.7.2

Physical Specifications

Analyzer

Length: 211 cm (83 in)

Width: 47 cm(18.5 in)

Height: 55 cm (21.5 in)

Weight: 74.5 kg (164 lb)

Optional Cryocooler Compressor

Length: 31 cm (12 in)

Width: 45 cm (18 in)

Height: 38 cm (15 in)

Weight: 32 kg (71 lb)

Power Requirements

Analyzer: 90-264 Vac, 47-63 Hz, 50 W (max) 30 W (typical)

Optional Heater: 90-264 Vac, 47-63 Hz, 150 W (max)

PC: 115/230 Vac, 50/60 Hz, 150 W

Optional Cryocooler Compressor: 100, 120, 220, or 240 Vac, 50/60 Hz, 500 W

Optional sample pump (RB0021-L): 115 Vac, 60 Hz, 950 W (other power options are available)