Dynamic parameters, Dynamic parameters 31 – Campbell Scientific TGA100 Trace Gas Analyzer Manual User Manual

Page 31

Advertising
background image

3.4.4

Dynamic Parameters

The next menu column, labeled “PARAMETERS”, provides access to the dynamic parameters, i.e. those that may be
changed in real time. The <up/down arrow> keys are used to scroll through the list of dynamic parameters (listed in

). If the Run/Edit mode is selected, the value of the selected dynamic parameter may be changed using the

<Home End PgUp PgDn + - / *> keys. Use the <Home End> keys for coarse adjustments, <Page Up Page Down>
keys for normal adjustments, the <+ -> keys for fine adjustments, and the </ *> keys for very fine adjustments. Number
pad and keyboard give the same results.

Table 3

Table 3. Dynamic Parameters

Some dynamic parameters may be controlled automatically. In this case, the value is displayed and updated in real time,
and an “A” will appear to the right of the value. Changing the value of a dynamic parameter that is being controlled will
disable the control function.

The dynamic parameters are a subset of the parameters stored in the parameter file and can also be edited through the
parameter change menu (see section 3.5). Table 3 lists the dynamic parameters by the abbreviated name shown in the
real time screen and gives the full parameter name shown in the parameter change menus (see section 3.5). Changes
made to these parameters in the Real Time Screen are saved in the parameter file.

Abbreviated Name Units Full Parameter Name

1

Laser Temp

K

Laser operating temperature (K)

2 DC

Current

mA

Laser DC current (mA)

3

Mod Current

mA

Laser Modulation current (mA)

4

High Current

mA

Laser High current offset (mA)

5

Zero Current

mA

Laser Zero current (mA)

6

SMP Gain

--

Sample detector gain

7

SMP Offset

--

Sample detector offset

8

SMP Det Temp

ºC

Sample detector operating temp (deg C)

9

REF Gain

--

Sample detector gain

10

REF Offset

--

Reference detector offset

11

REF Det Temp

ºC

Reference detector operating temp (deg C)

12

Std Dev Time

sec

Mean, StdDev time frame (sec)

13

Graph3 Range

--

Graph 3 range

14

SMP Det Lin

--

Sample detector linearity coeff

15*

RampB Offset

mA

Ramp B offset current (mA)

16*

RampB Mod

mA

Ramp B Modulation Current (mA)

17*

RampB High

mA

Ramp B High current (mA)

18*

SmpDetLin B

--

Ramp B Smp detector linearity coeff

* Dynamic parameters 15 through 18 are available only in dual ramp mode.