Figure 5-5, Figure 5-5. gradient measurement at four sites – Campbell Scientific TGA100 Trace Gas Analyzer Manual User Manual

Page 73

Advertising
background image


that site’s gradient valve assembly to a corresponding inlet on the site selection assembly located near the TGA100. The
site selection assembly consists of four nonlatching three-way valves, one for each site. The outlets of the four valves
are connected to a manifold that is connected to the TGA100 sample inlet. The TGA100 controls the site selection
system using timing parameters supplied by the user. Normally each site is measured for 15 to 60 minutes before
switching to the next site.

Purge Dryer

Sample Dryer

Site Selection Control Cable

Sample

Bypass

Dryer Purge

Sample

Gradient

Intake

Filter

Needle Valve

Gradient Valve

Site 2

Site 3

Site 4

Site 1

Assemblies

Valve
Assembly

Control Cable

Sample Tubing

Gradient Valve
Control Cables

From Each Site

To Each Site

Site Selection
Sampling
System

Figure 5-5. Gradient Measurement at Four Sites

To configure the TGA100 for this example site selection assembly:

1) Connect digital output bit 2 to site valve 1, bit 3 to site valve 2, bit 4 to site valve 3, and bit 5 to site valve 4.

73