Omitted data count, Omitted data count 62, Figure 4-12. dual ramp scan sequence – Campbell Scientific TGA100 Trace Gas Analyzer Manual User Manual

Page 62: Figure 4-13. adjustment of omitted data counts

Advertising
background image

Laser
Current

CO

2

13

CO

2

12

Ramp B Offset

Ramp A

Ramp B

Detector
Response

Figure 4-12. Dual Ramp Scan Sequence

To set the ramp B high current, first set the (ramp A) high current as described in section 4.4.3. Then repeat the process
for the ramp B high current. When evaluating the ramp B high current setting,, observe the ramp B line instead of the
ramp A line, and use both the ramp A line lock and the ramp B line lock to center the absorption lines. Count steps
while adjusting the DC current, as was done for the ramp A high current (not the ramp B offset current).

4.4.4

Omitted Data Count

Some additional data must be omitted from the concentration calculation to allow the laser temperature to stabilize fully
after the zero and high current phases of the scan. The omitted data count parameter specifies how many scan points to
omit It may have a value from 4 to 20 counts, where each count represents a 20

µs interval, to give a total duration of

omitted data of 80 to 400

µs. To set this parameter, look at the reference detector transmittance in Magnified mode. The

leftmost of the three vertical dotted lines on the display marks the end of the omitted data counts, i.e., the start of the
data used to calculate concentration. Increase the omitted data counts to move this line to the right; decrease it to move
it to the left. Set the omitted data counts to avoid a transient at the start of the points used in the concentration
calculations. When in doubt it is usually better to omit a few extra points. Figure 4-13 illustrates how to set this
parameter.

Too few omit counts

Correct omit counts

Drops at

left edge

Flat at

left edge

Figure 4-13. Adjustment of Omitted Data Counts

For dual ramp mode, the omitted data count parameter is applied to ramp A and ramp B, and it should generally be set
to its maximum value.