3.5.3

Analog Output Screen

The Analog Output screen allows the user to configure the analog output channels (see section 0). Use the
<up/down/right/left arrow> keys, or the <TAB> and <shift-TAB> keys to move to the field to be changed. To
change which data will be output, highlight the desired channel, and type <enter>. A new menu will appear that will
show the options available for output. These options are the same as for the real time graphs and for output to the 10 Hz
data file, and are listed in Appendix A. Use the <up/down/right/left arrow> keys, the <Pg Up / Pg Down> keys, and
the <TAB> and <shift-Tab> keys to select the desired option, and then hit the <enter> key.

If the TGA is running, the current value of the parameter and the corresponding analog output voltage will be
displayed, as shown in

.

Figure 3-6

Figure 3-6. Example Analog Output Screen