Figure 5-7 – Campbell Scientific TGA100 Trace Gas Analyzer Manual User Manual

Page 77

Advertising
background image

100

200

300

400

500

600

700

800

900

1000

0

10

20

30

40

50

60

70

80

90

100

Tube Length (m)

Pressure (mB)

1 slpm

2 slpm

3 slpm

5 slpm

7 slpm

0

2

4

6

8

10

12

0

10

20

30

40

50

60

70

80

90

100

Tube Length (m)

Time (s)

1 slpm

2 slpm

3 slpm

5 slpm

7 slpm

Figure 5-7. Pressure Drop and Travel Time for 0.125" (3.2 mm) ID Tubing

100

200

300

400

500

600

700

800

900

1000

0

20

40

60

80

100

120

140

160

180

200

Tube Length (m)

Pressure (mB)

2 slpm

3 slpm

5 slpm

7 slpm

11 slpm

0

5

10

15

20

25

30

0

20

40

60

80

100

120

140

160

180

200

Tube Length (m)

Time (s)

2 slpm

3 slpm

5 slpm

7 slpm

11 slpm

Figure 5-8. Pressure Drop and Travel Time for 0.170" (4.3 mm) ID Tubing

100

200

300

400

500

600

700

800

900

1000

0

50

100

150

200

250

300

350

400

Tube Length (m)

Pressure (mB)

3 slpm

5 slpm

7 slpm

11 slpm

16 slpm

0

10

20

30

40

50

60

0

50

100

150

200

250

300

350

400

Tube Length (m)

Time (s)

3 slpm

5 slpm

7 slpm

11 slpm

16 slpm

Figure 5-9. Pressure Drop and Travel Time for 0.250" (6.4 mm) ID Tubing

77