Error messages, Displaying the scpi error queue, Error list – Agilent Technologies System DC N5700 User Manual

Page 118

Advertising
background image

Appendix C Service

118

Series N5700 User’s Guide

Error Messages

Displaying the SCPI error queue

The entire error queue is read, then emptied, using the following
command: SYST:ERR?

Error List

The following table documents the various error messages that the
power supply supports:

Error Device-dependent

Errors

(these errors set Standard Event Status register bit #3)

0

No error
This is the response to the ERR? query when there are no errors.

100

Too many channels
You have specified more channels than are installed in the mainframe.

101

Calibration state is off
Calibration is not enabled. The instrument will not accept calibration commands.

102

Calibration password is incorrect
The calibration password is incorrect.

104

Bad sequence of calibration commands
Calibration commands have not been entered in the proper sequence.

114

CAL:DATE must be yyyy/mm/dd
The calibration date must be entered in the numeric format yyyy=year, mm=month, dd=date

203

Compatibility function not implemented
The requested compatibility function is not available.

204

NVRAM checksum error
A checksum error has occurred in the instrument’s nonvolatile random access memory.

205

NVRAM full
The nonvolatile random access memory of the instrument is full.

206

File not found
The internal calibration file or the internal channel attribute file was not found in NVRAM.

209

Output communications failure
A hardware failure has occurred on the power supply.

302

Option not installed
The option that is programmed by this command is not installed.

351

VOLT setting conflicts with VOLT:PROT setting
Attempted to program the voltage above the over-voltage protection setting.

352

VOLT:PROT setting conflicts with VOLT setting
Attempted to set the over-voltage protection below the voltage setting.

353

VOLT setting conflicts with VOLT:LIM:LOW setting
Attempted to program the voltage below the under-voltage limit setting.

354

VOLT:LIM:LOW setting conflicts with VOLT setting
Attempted to set the under-voltage limit above the voltage setting